OPTICAL BUILD CARD / A-E1 • millimetres; world coordinates Every dimension marked "nominal" is a design target. Record the as-built position, axis sign, device IDs and the method used to measure each quantity. source centre (120,+240,120) ↙ red light source side Y > 0 outer mouth at +25 shared datum exit mouth at Y = 0, centre (0,0,120) receiver side Y < 0 active plane Y = −70, normal parallel to +Y direct-field centre (−35.053,−70,120) X = lateral; Z = up rectangle's 50 mm side along X 1 Source head Assemble on an adjustable axis: red star on heat spreader, opal diffuser (Edmund 34-473, 25 mm OD × 1.25 mm nominal), illuminated circular metal stop, then lens (Edmund 15-542, 75 mm OD, 72 mm clear). Put diffuser very close behind the stop with no rim in its opening. Extend the illuminated diffuser beyond the stop; adjust LED–diffuser gap, polish burrs and record the selected distance. Allow translation for focus and transverse centring. Orient the condenser's planar face toward the small focal stop and its aspheric face toward the approximately collimated beam. Contact the rim with compliant pads, allow for actual glass OD (74.5–75 mm), and avoid hard clamping. The lens centre is nominally 30 mm thick (±0.5), edge about 3 mm. Measure the retaining ring and housing ID along the tilted axis. Target an unobstructed 72 mm pupil; 65.96 mm is the geometrical requirement at the stated position for rays from the whole ideal exit reference. The 6.04 mm diameter margin is only about 3.02 mm per side. The old 50 mm mock-up cuts the field. Cut interchangeable starting stops around 3.0, 3.5 and 4.0 mm; measure their finished diameter and edge profile. For EFL 50 mm, 2f tan 2° = 3.492 mm. Use the vendor's 30.5 mm BFL as an initial focal reference from the planar surface; mount the stop on an axial slide spanning at least 20–45 mm from that face for adjustment. EFL tolerance ±7% means even a perfect 3.492 mm stop need not give the target half-angle. Actual aberration, tilt, diffraction and source fill also matter. Locate best focus with the *assembled* unit. Fix U4 at a baffled side port sampling diffuse light before the stop, out of the transmitted bundle. Mark its position and check its signal with each specimen/mask installed, removed and shuttered. A stable monitor does not establish a stable beam profile. If U4 saturates, reduce the monitor port or add its own attenuator, then characterize it relative to the beam again. 2 Characterize light before machining the final pair Sight pupil and exit centres at the angle above. Verify source axis and opening with independent scales. Scan the bare exit plane for spatial brightness and record the field at the centre, near the square and rectangle corners and at the long rectangle ends. Stay inside the real clear areas and record coordinates. A spatial map alone cannot verify the 2° angular cone. Workshop option for directional sampling: make two thin blackened pinholes of about 0.5 mm diameter, 120 mm apart, in a baffled tube; place the entrance hole on a two-axis pivot at the chosen exit-plane point. Place U3 immediately behind the rear stop, shield electronics and ambient light. A relative ray within atan(0.5/120) ≈ 0.239° of the axis may pass the two ideal equal holes. This is a finite-response half-support estimate, not angular precision. For 625 nm light a 0.5 mm hole also has a diffraction scale of ~0.087° (1.22λ/D). Test a reference beam or borrow a lab angular camera/goniometer to measure the *actual* instrument response and angle zero. At each position sweep azimuth and elevation in the local tangent plane around the nominal source direction, e.g. ±4°, 0.5° rough steps then 0.25° near edges; widen while measurable tail remains. Measure darks and repeat sweeps. Check that spatial positioning and pivot repeat through the sweep. Integrate angular cells with their actual solid angles and source-to-mouth cosine for the prediction. Record receiving jig response and relative sensitivity at each angle; avoid adding a second cosine if already built into the calibration. No assumed angular hard cutoff. Estimate throughput BEFORE spending time on a sweep. For a uniform 2° cone, equal 0.5 mm holes have area 0.19635 mm² each. The exit aperture subtends ~1.36e−5 sr at 120 mm versus ~3.83e−3 sr for the full cone. Relative to a bare 2.29 mm square head, the fraction near beam centre is of order (0.19635/5.244) × (1.36e−5/0.00383) ≈ 1.3e−4, ignoring tilt and losses. Thus a 1 V bright open-head signal may yield only ~0.13 mV or two ADS counts above its dark pedestal at our gain. This is a design estimate, not a measured noise floor. The two-hole fixture is a coarse aiming/feasibility option. Run blocked, centre and off-axis repeats first. Do not rely on it for a qualified angular map unless measured uncertainty supports the required discrimination. Preferred qualification loan: a lab goniophotometer or a small RAW-capable linear camera with a focal-plane lens and known angular response. For example, place a 2–3 mm sampling iris at the chosen mouth point, align a nominal 35 mm focal-length lens immediately behind it along the central ray, and image its back focal plane on the sensor. No aperture specimen installed. Ray angle maps approximately as radius = f tan(angle): ±2° -> ±1.22 mm; ±4° -> ±2.45 mm. Require ≥4.9 mm of usable image area in BOTH directions for that field or change focal length/scan overlapping pointings. Provide mechanical access for focus and both axis signs. Measure iris position, angular zero, pixel-to-angle mapping, point spread, vignetting and linearity with known directions and a characterized input; document finite iris spatial averaging. Capture linear raw counts with fixed exposure/gain, dark/flat frames and unsaturated core; bracket exposures for spill with overlap and measured linear response. Repeat at multiple mouth positions and temperatures. Aim for measurable residuals after model sensitivity analysis. If the accessories, operator or calibration cannot be borrowed/quoted, retain the prototype and label its angular characterization unresolved. Include paid lab time in the cost sheet before claiming a complete qualified instrument. 3 Receiver head and range Stay at Y = −70 with the die face parallel to the mouth. Survey the optical centre on the sensor itself and use that plane as datum. Nominal direct full-reference footprints: rectangle 56.11 × 17.97, square 31.11 × 30.47, about (−35.05,120) in X,Z. For a full 100 × 75 trial field, plan X = [−85.05,+14.95], Z = [82.5,157.5]. Confirm actual travel after end switches and soft limits. The square and rectangle use the same installed/stage datum. These bounds include direct rays under the design cone. They do not bound all scatter. Start with coarse reference maps on both mouths, then expand the scan at each edge. Integrate the additional ring; record change, noise and maximum accessible extent. A dark perimeter is a diagnostic, not a bound on all missed wide-angle energy. Borrow a characterized collector or compare geometries if the residual cannot be bounded to the accepted budget. Repeat with fine and half-pitch-shifted grids and remounts. If changing receiver plane, remeasure coverage and position rather than using the original bounds. TI's active area is 2.29 × 2.29 mm. Verify the *housing* does not vignette. For a light tunnel of axial length t, the oblique footprint translates by up to ~0.545t in X and ~0.040t in Z for the design cone. A centred, unshifted rectangular window thus needs at least 2.29 + 2×0.545t width and 2.29 + 2×0.040t height plus placement margin at its outer plane. At t=2 mm the geometric minima are 4.47 × 2.45 mm before tolerances. Prefer a flush, wider relieved window. Survey angular response of the complete head including package/window, with a stable beam and fixed incident flux. Extend angles or carry an uncertainty for scattered light; its angular distribution can differ between specimen and reference. The same head does not guarantee cancellation. 4 Acquisition and analysis transfer Acquire both-head darks with both windows covered, plus shuttered specimen- plane background with the source on and the monitor lit. Repeat before/after maps and after drift; track temperature, LED current and beam profile. A shutter should sample unrelated background without introducing a new bright surface. Treat real specimen-scattered power as part of the observable. A bracket at a stable point is monitor (m1,t1), head (h,t), monitor (m2,t2). Use individually dark-subtracted monitor values, interpolate at t between t1 and t2, and divide background-corrected head by that estimate. Use the time midpoints of the recorded conversion windows and preserve raw windows. Reject and label unstable/near-zero denominators. Subtract a normalized shutter background consistently; do not subtract the electrical dark twice. Time budget: 32 brackets/point need 96 conversions. At 128 conversions/s the theoretical lower bound is 0.75 s/point. A 101×76-point, 1 mm map is ≥96 min; a 201×151-point, 0.5 mm map is ≥6.32 h before move/settle, communications, USB reset, references and remounts. The spot CLI uses a transient USB command; the raster runner in RUN_A_RASTER.txt holds both device connections open. Actual overhead from command, controller, conversion, settling and file I/O must be timed. A single 10-point timing run, a long stationary stability log and a home-return repeat reveal whether an efficient continuous host link, less bracketing, a different pitch or splitting maps into referenced blocks is necessary. Set choices from commissioning data before final comparisons. Separate spatial settling from ADC conversion. At a bright/dark spatial edge, move back to a fixed point from both directions; take bursts immediately and after candidate waits such as 0.1, 0.3, 1 and 3 s with motors in their final hold state. Compare early/late corrected values, repeatability and position readback against a late steady window. Repeat on a steep gradient: a flat uniform field hides position errors. With a repeatedly sampled reference level, propagate the observed timing contribution to T before choosing the shortest acceptable settle. This procedure determines a number specific to the actual rig and motion profile; the electronic timeout is unrelated. For a rectilinear map integrate normalized readings by physical area weights (endpoint trapezoid factors if appropriate). For a spatially invariant uniform active area divide by the effective head area to recover total flux; a common constant cancels in a ratio. Characterize angular and spatial response separately where it changes. Do not count overlapping spots as independent flux samples or clip negative dark-corrected tails to zero. For each specimen use its own thin same-mouth mask on the exit datum. Let eta_mask=P(mask)/P(ideal matched open mouth) from the actual mask/angle field or a characterized comparison. T = eta_mask × M(recess)/M(mask); uncertainty in eta is shared wherever the same reference is reused. The original 0.5 mm mask estimate would bias uncorrected T by ~0.00513 (square), ~0.00413 (rectangle). Bracket recess scans by references, balance specimen order, include remounts and the 180° rectangular roll. Record dimensions, reference exposure, mask thickness/profile, tape/paint batch, wall leaks, pre/post profiles and control deviations. Freeze the interpretive protocol before production data. Keep raw counts, corrected maps and prediction files distinct. Engineering checks end when the probe rests and returns, the spatial field is covered, state IDs match the records and one reference–recess–reference trial is repeatable with no overload or lost records. Scientific sign-off is separate. Use measured bounds or a correlated uncertainty model for detector linearity, temporal drift, aperture response, angular map, collection/truncation, mask transmission, geometry and wall scatter. The inherited planning half-width 0.005 in T apportions 0.0005 to dark/readout/repeatability, 0.001 to drift/ nonlinearity, 0.002 to collection/response, 0.001 to reference and 0.0005 reserve. These are allocations only. Evaluate measurement and predicted intervals independently then apply the predeclared comparison (traveler p.3). Unaffordable or unresolved calibration leaves a bounded prototype result. Source/circuit specifications cannot determine scientific accuracy in advance. Vendor sources: https://www.edmundoptics.com/p/75mm-diameter-x-50mm-fl-mgfsub2sub-coated-aspheric-condenser-lens/43312/ https://www.edmundoptics.com/p/25mm-dia-white-diffusing-glass/3837/ https://www.ti.com/lit/ds/symlink/opt101.pdf Existing numerical design inputs: measurement_design/20261026/LAYOUT_CALCULATIONS.json Replication sequence: ../README.md. Revision 9 model: ../prediction/protocol.txt.